AX-UL-LEP-1x2-2 / Ultrasonic Level Transmitter, 0-14pH, Plastic Housing BSP 1½ PP, 4-20mA POA Qty
AX-UL-LER-1x2-2 / Ultrasonic Level Transmitter, ±1000 mV, Plastic Housing BSP 1½ PP, 4-20mA POA Qty
AX-UL-LED-1x2-2 / Ultrasonic Level Transmitter, 0 – 20 ppm v.0 – 10 ppm, Plastic Housing BSP 1½ PP, 4-20mA POA Qty
AX-UL-SAP300 / Display Unit for Ultrasonic Level Transmitter “AX-UL-L” series POA Qty
 <<  |  <  | OVERVIEW | SPECIFICATION | DATASHEETS        

AX-ANA-L Liquid Analytical Transmitters, Designed for measuring pH, Redox (ORP), and Dissolved Oxygen (DO) in liquids.pH: Detects the voltage difference between measuring and reference electrodes to determine acidity/alkalinity. Provides temperature-compensated pH (25°C) with proportional output. Periodic calibration required.ORP: Measures oxidation–reduction potential via electrode voltage on a platinum surface. Outputs a signal proportional to the oxidising/reducing balance in the medium.DO: Uses a membrane-covered sensor generating a signal propor-tional to dissolved oxygen concentration. Electronics provide tem-perature-compensated DO (25°C) for monitoring and control.
Product Features
Compact and Integrated Versions.
4-20mA, HART, Relay Output
Separated versions up to 10m
IP67 / IP68 protection
Measurement range: pH: 0-14, ORP: ±1000 mV, DO: 0-20 ppm
Plug-in display unit

Measuring Range
pH
ORP
±1000 mV, Reserve (±200 mV)
DO
0 – 20 ppm v. 0 – 10 ppm, Reserve (20%)
System
2 Wire
Accuracy
pH & ORP (0.1% of the measured value ±1 digit ±0.01% / °C)
DO (0.5% of the measured value ±1 digit ±0.01% / °C)

Measuring Cycle
300 msec, on display
Electrode Input
pH & ORP
DO
DO sensor input
Output
4–20 mA, (3.9 - 20.5mA), Rtmax = 1200Ohm galvanic isolated, protection against surge transients
Temperature Measuring (semiconductive sensor)

Range
-50 to 130°C, Accuracy
Ambient Temperature

Aluminium housing
-30°C to +70°C, Plastic housing
Power Supply
12-36 V DC / 48 mW-720 mW, Galvanic Isolated, Protection against surge transients.
Housing
Compact type
Electrical Connection
Compact type
O7-13 mm wire cross section
0.5-1.5mm2 (shielded cable is recommended), + 2 x NPT 1/2" internal thread for cable
protective pipe Integrated type
6x0.5 mm2 shielded cable, O6 mm x 5 m standard (up to max. 30 m cable length)
Ingress Protection
IP67
Electrical Protection
Class III. electric shock protection
Weight
<7 kg (15.4 lb)
Country Of Origin
Hungary

AX-ANA-L.pdf / Analytical pH, ORP, and DO Transmitter