datasheetdetails
|
<< | < | AX-ANA-LED - Dissolved Oxygen Sensor
|
|
The AX-ANA-LED-xxx is an Analytical sensor for the measurement of the quantity of dissolved oxygen gas in the liquid. in either ppm or mg/l. The sensor with its oxygen permeable membrane is immersed in the liquid and provides an electronic signal proportional to the oxygen concentration There is the option of a relay and / or an LCD Display module. |
Product Features
| • Measuring range 0-10 or 0-20ppm • Easy to set-up
| • Temperature compensated • Variety of probes available
|
Measuring range - LED-111
| 0-10ppm
| Measruing range - LED121
| 0-20ppm
| Resolution
| 0.01ppm (internal resolution 0.005ppm)
| Linearity
| +/- 0.05 ppm
| Accuracy
| 0.5% of the measured value +/- 1 digit +/- 0.01% /Deg C
| Measuring Rate
| 300msec, on the display (refreshing rate) 1 sec
| Process Temperature
| 0 +50 degC
| Power Supply
| 12 to 36Vdc 720mW galvanically isolated, protected against surge transients
| Output - LED-1X1-2
| 4-20mA @1200ohms
| Output - LED-1X1-4
| HART & 4-20mA
| Output - LED-1X1-6
| 4-20mA @1200ohms EeX
| Output - LED-1X1-8
| HART & 4-20mA EeX
| Output - LED-1X1-R
| 4-20mA & Relay 1 x SPDT 1A @ 30Vdc
| Output - LED-1X1-H
| HART & 4-20mA & Relay 1 x SPDT 1A@30Vdc
| Display (option) SAP300
| Full featured programming LCD Display
| Intrinsically Safe Class
| ATEX II 1 G Exia B T6 Ga
| Cable Length
| 5m provided
| Weatherproof Rating Probe
| IP68 Electronic IP67
| Materials Housing
| Plastic (aluminium available at additional cost)
| Process Connection
| 1 1/2" BSP Polyproylene as standard ( PVDF option at additional cost) NPT option
| Electromagnetic Compatability
| EN61326 Class B
| Country of Origin
| EU
|
AX-ANA-LED.pdf / Analytical sensor- Dissolved oxygen transmitter |  |
|
|
|